4VS Monitor

Very much a monitor,

Very much a Spengler 

  • Type

    Vital signs monitor

    • Blood pressure
    • SPO2
    • Temperature
    • ECG
  • Blood pressure

    Oscillometric method

    3 modes

    • Adult patients
    • Pediatric patients
    • Neonatal patients

    Measurement range

    • Adult : Sys 40-60 mmHg / DIA 20-200 mmHg
    • Pediatric : Sys 40-160 mmHg / DIA 20-120 mmHg
    • Neonatal :  Sys 40-130 mmHg / DIA 20-100 mmHg

    Resolution

    • 1 mmHg

    Accuracy

    • +-3 %

    Overpressure limit

    • Adult : 290 mmHg
    • Pediatric : 220 mmHg
    • Neonatal : 150 mmHg

     

  • SPO2

    Method

    • Dual wavelength infrared wave

    Measurement

    • From 0% to 100%

    Alarm configuration range

    • From 70 % to 100 %

    Resolution

    • 1 %

    Accuracy

    • Adult et pediatric : 2 % between 70 % and 100 %
    • Neonatal : 3 % between 70 % and 100 %
  • Temperature

    Measurement mode

    • Thermal

    Measurement and alarm range

    • From 0°C to 50°C (32°F – 122°F)

    Resolution

    • 0.1°C

    Accuracy

    • Between 25°C between 45°C :  0.1°C
    • Other : 0.2°C

    Average measuring time

    • < 60 seconds

    Data units

    • C° or F°
  • ECG

    Method

    • 3 diversions (RA, LA , LL) / 5 diversions (RA, RL, LA, LL, V)

    Scanning speed

    • 12.5 mm/s, 25 mm/s, 50 mm/s

    Heart rate range

    • Adult : 20 ~ 300 bpm
    • Neonatal / children : 20 ~ 350 bpm

    Resolution

    • 1 bpm

    Accuracy

    • From 20 to 200 Bpm : 5 %
    • From 201 to 350 : 10 %

    Alarm configuration

    • Adjustment range : 20 ~ 350 bpm

    Alarm method

    • Alarm with sounds and lights, record data during alarm for examination
  • Terms and conditions of use

    Temperature

    • Use : From 5°C to 40°C
    • Storage : From -40°C to 55°C

    Relative humidity

    • Use : Below 95% non-condensing
    • Storage : Below 95% non-condensing

    Atmospheric pressure

    • Use : From 70 kPa to 106 kPa
    • Storage : From 16.5 kPa to 106 kPa

     

  • Size and weight

    Dimensions

    • Device : 180 mm × 250 mm × 180 mm
    • Packaging : 310 mm × 320 mm × 235 mm

    Weight

    • Device : 2 kg (including batteries)
    • Packaging : 3.7 kg
  • Power

    Power supply

    • 100 ~ 240 VAC (± 10 %), 50 / 60 Hz (± 3 Hz), 2 A, 40 VA Max.

    Internal battery input

    • 6.4 V ~ 8.4 V, 2200 mAh, reloadable lithium battery

Spengler Warranty : 2 years for the monitor (under normal operating conditions)

Document(s) to download